As a core component of industrial gas delivery systems, the sealing performance of stainless steel pressure reducers directly impacts operational safety and equipment lifespan. However, factors such as mechanical wear, material corrosion, and improper operation frequently lead to air leakage issues. This article systematically outlines key safety management points for stainless steel pressure reducers from three dimensions: causes of leakage, detection methods, and maintenance strategies.

Mechanical seal failure is the primary cause of leaks.
For instance, threaded connection loosening often results from installation without a torque wrench, leading to high-pressure gas escaping at joints.
Gasket aging is linked to rubber material oxidation and brittleness in oxygen environments; one laboratory experienced an oxygen system leak due to expired gaskets.
Additionally, damage to valve components is common. Scratches exceeding 0.1mm on valve seats can allow gas pressure to breach the sealing surface, creating leakage pathways.
Material corrosion risks are particularly pronounced in corrosive gas environments.
While 316L stainless steel offers corrosion resistance, it remains susceptible to pitting corrosion in media like chlorine and ammonia.
A semiconductor company failed to install filters on chlorine pipelines, causing particulate contaminants to jam the throttle port of a pressure reducer and rupture the diaphragm, leading to leakage.
Operational errors also warrant attention.
Rapid valve opening induces adiabatic compression heating, causing rubber packing temperatures to surge above 200°C. This resulted in a combustion incident involving an acetylene pressure reducer at a shipyard.
Furthermore, grease contamination poses a lethal hazard in oxygen systems. At a machinery plant, an operator failed to wear oil-free gloves while installing an oxygen pressure reducer, causing high-pressure oxygen to react with grease and trigger an explosion.
Sensory detection is suitable for preliminary screening.
The sense of smell can detect leaks of gas with added odorants;
hearing can detect a “hissing” sound at leak rates as low as 0.5g/s;
while touch can detect gas flow vibrations by lightly brushing the back of the hand over suspicious areas.
Instrumental detection enables precise localization.
Soap solution testing involves applying a 10% sodium dodecyl sulfate solution and observing bubble formation to pinpoint leaks;
electronic leak detectors use semiconductor sensors to detect gas concentrations at the 0.1 ppm level with response times <3 seconds;
Helium mass spectrometer leak detectors operate in vacuum mode, displaying leak rates as low as 1×10⁻¹² mbar·l/s, suitable for high-precision applications like semiconductors.
Pressure testing verifies sealing integrity through system pressure changes.
Pressure retention tests require no more than a 2% pressure drop from initial value within 10 minutes after cylinder valve closure.
Differential pressure testing uses dual pressure gauges to distinguish leaks on high-pressure versus low-pressure sides.
Proper installation is fundamental to leak prevention.
Ensure the installation environment is dry, well-ventilated, and distant from fire sources and corrosive substances.
When connecting cylinders to pressure regulators, torque values must comply with manufacturer specifications, typically 30-40 N·m.
Install filters in piping systems to intercept particulate contaminants, preventing throttle orifice blockage or diaphragm damage.
Operational maintenance requires regular inspection and cleaning.
Monthly purging with dry compressed air removes internal dust, while annual professional disassembly cleaning is mandatory.
If freezing occurs, thaw only with water below 40°C; flame heating is strictly prohibited.
Monitor pressure gauge readings during operation; immediate shutdown and inspection are required if pressure rises persistently or exhibits abnormal fluctuations.
Storage management requires categorized storage and periodic inspection.
Maintain a minimum distance of 5 meters between oxygen and acetylene pressure regulators, and isolate them from greasy substances.
Store in environments with relative humidity controlled between 30% and 60%, avoiding impacts or vibrations.
Conduct comprehensive inspections every three years to verify pressure gauge accuracy and replace aged seals.
Safety management of stainless steel pressure regulators requires establishing a comprehensive “prevention-detection-response” control system. By strictly adhering to installation standards, implementing regular inspection protocols, and utilizing intelligent maintenance technologies, leakage rates can be effectively controlled below 0.05%, providing robust assurance for the safe operation of industrial gas systems.